SPECIFICATIONS |
|
System Voltage (50Hz/60Hz) |
550V, 380V, 240V |
Current Settings |
2-999mAac/50Hz |
Current Selection |
Knob |
Phase Start Selection Referenced to Earth |
0 degree and 180 degree |
Over-Temperature Protection |
Yes |
Wiring Correctness Indication |
Yes |
Phase Polarity Trip Indicator |
Yes |
Operating Voltage (L-E) |
110 Vac to 317 Vac |
Timer Resolution |
1mS (max time=19.999s) |
Timer Accuracy |
+/-1% +/-1mS |
Current Accuracy |
+/- 1% +/-1mA |
Acurrent Resolution |
1mA |
Voltmeter Accuracy (50Hz) |
50-350 Vac = +/- 2% +/- 1V, 350-450 Vac =+/- 5% +/- 1V |
Voltmeter Resolution |
1V |
Operating Temperature |
-5 degree C to 45 degree C |
Storage Temperature |
-10 degree C to 85 degrcc C |
Power Source |
1.5V (AA) x 8 batteries, Bat OK Led Vbat > 7.5V |
Dimensions |
170(L) x 165(W) x 92(D) mm |
Weight |
Approx. 1300g (battery included) |
Accessories |
Test leads (TEL-1813), shoulder belt (BET-1800), instruction namual, batteries |